General
The principle of pHmeasurement is quite simple. All one needs is aglass electrodeconstructed of pH sensitive glass and a reference (potential) that remains stable under all conditions.This directly shows the weak point inthe measuring loop. As the reference needs an ‘open’ circuit to the process, a stable reference potential is sometimes hard to keep. Mostreference electrodes have aporous diaphragm to realize this open
connection to the process. Several materials are being used like ceramic, Zirconia, PTFE and wood. The porous diaphragm can
get clogged or high diffusion potentials may build up.Po***ning of the reference is another cause for unstable reference potentials.
A reference electrodecan be po***ned by the penetration of
the process liquid or by diffusion of components of the process medium through the diaphragm. Thestable reference potential which is formed by a Ag/AgCl referencesystem in aKCl solution is shifted if either gets effected.E.g. silver chloride +sulfide will form the insoluble silver sulfide. These particles give deposits, which canblock the diaphragm from inside the reference electrode.Likewise the chloride concentration can be altered and therefore thereference potentialis altered. The latter can be corrected by re-calibration. Po***ningof the AgCl referencesystem is irreversible and the reference electrodeneeds to be replaced.
Maintenance
For accuratepH measurement, regular (even frequent)
maintenance is required.This knowledge is decisive for the
selection of fittings and electrodes of the pHmeasuring loop. One needs to keep in mind that the electrode(s) must remain accessible and exchangeable. Once the right fitting has been selected, the process of selecting the right electrode(s) must take place.
Glass electrode
The SM21-AG4 will suit the great majority of the pH applications as this electrode is constructed in such a way that its sensitivity and strength are optimized. When sensitivity isrequired more than strength than the SM21-AG2should be selected.
For toughindustrial applications (including auto-cleaning) thedome shaped electrode is the best selection. ‘L’-glass should be selected when the electrode is used at higher temperatures (>70ºC) and/or in applications that need a glass type that isresistant to chemical attack.
Reference electrode
Reference electrodes come in flowing- and non-flowing types.Flowing types such as the SR20-AC52, AC32 with a ceramic diaphragm are used in processes that have low ionic strength such as (ultra) pure water. The SR20-AS52 is aflowing type electrode with asleeve junction.Due to the high electrolyte flow rate, the sensor is suitable for UPW application as well as in highly polluting applications. The electrolyte flow towards the processtends to keep the junction clean from clogging and po***ning.
For all other applications the SR20-AP24 is preferred.When the process cannot withstand KCl, the SR20-AP26 should be